Catalog Number 18471136S
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/01/2014
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.Device not available to stryker.
 
Event or Problem Description
The manufacturer became aware of a post market clinical follow-up (pmcf) from (b)(6).The title of this report is ¿analysis of the t2 recon nail used with the antegrade locking mode¿ which was released in june 2018 and is associated with the t2 recon nailing system.Within that report, postoperative complications/ adverse events were reported, which occurred between january 2014 and november 2017.Based on the available device and patient information and a review of complaint handling database it was revealed that the events have not been reported by the hospital or by the author of the report, therefore 15 complaints were initiated retrospectively for different adverse events mentioned in the journal.This product inquiry addresses implant breakage.Two (case id 62) out of 3 cases.The pmcfr states: ¿in 3 cases (case 20,62 and 71) breakage of the material occurred.The breakage was situated on the locking screws and never on the nail itself.In the second case (case 62) it was a transverse fracture (32-a) in a heavy 70 years lady.She is 1,63m in height for a weight of (b)(6).The breakage of all three locking screws was observed.In fact a septic nonunion was observed and a revision surgery was performed 142 days after the initial surgery.All the broken screws were removed (18965050s-k05d869; 18965050s-k0af9e2; 18965042s-k074ffc), the nail was also removed (18471136s-k0dadeb) the medullary canal was reamed and a t2 a/r nail was then used.Antibiotics were used during 4 months.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECONSTRUCTION NAIL R2.0, TI, RIGHT T2 RECON Ø11X360 MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
MDR Report Key7710515
Report Number0009610622-2018-00373
Device Sequence Number789926
Product Code HSB
UDI-Device Identifier07613153316961
UDI-Public07613153316961
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K032898
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number18471136S
Device Lot NumberK0DADEB
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2018
Initial Report FDA Received Date07/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight96
-
-