BOSTON SCIENTIFIC - GALWAY PROMUS PREMIER¿; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number H7493925216250
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Arrest (1762)
Date of Event 06/25/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is a combination product. the device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications. the investigation conclusion is user / use error as there was an act or omission of an act that resulted in a different medical device response than intended by the manufacturer or expected by the user.  (b)(4).
 
Event or Problem Description
Same case as mdr id: 2134265-2018-07108, 2134265-2018-07109, 2134265-2018-07110.It was reported that patient ¿coded¿.The 90% stenosed target lesion was located in the left anterior descending (lad) artery.The patient also had a chronic total occlusion in the right lad as well as significant disease in the circumflex and in the obtuse marginals.There were a total of 4 bsc stents deployed in the lad and only one was expired which was the 16 x 2.50 promus premier¿ stent.The 3 other stents were one promus otw and 2 synergy mr stents.The expired stent was discovered after the case, when the devices used were being consolidated.The procedure was completed and the patient was fine.However, 3 days post-procedure, the patient coded.A temporary pacemaker was implanted to control the patient's heart rhythm.No further patient complications were reported and the patient was recovering.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER¿
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7710617
Report Number2134265-2018-06359
Device Sequence Number2580288
Product Code NIQ
UDI-Device Identifier08714729828860
UDI-Public08714729828860
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P110010
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/16/2018
Device Model NumberH7493925216250
Device Catalogue Number39252-1625
Device Lot Number19885264
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2018
Initial Report FDA Received Date07/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-