Model Number N/A
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D6: implanted date: device was not implanted d7: explanted date: device was not explanted the actual device has not been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.For this reason, evaluation code 20 has been referenced in the conclusions section of h6.The production lot number was not provided by the user facility, which prevented a meaningful review of the device history record, the shipping inspection record or the complaint file.Terumo medical products (tmp) (importer) registration no.2243441 is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.9681834.Exemption number e2015022.
 
Event or Problem Description
The user facility reported that a doctor was performing a radial procedure and the glide sheath slender had the distal shaft break off and had to be removed from the patients radial artery.Additional information was received on 7/10/18.There was no medical or surgical intervention required.There was another catheter used.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to correct the catalog no.That was reported in the initial report.Provide the device return date.State that this report has been deemed not fda reportable.
 
Additional Manufacturer Narrative
This report is being submitted to correct the device return date that was submitted in the follow up no.2 report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESHEATH SLENDER
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, shizuoka 418
JA  418
MDR Report Key7710710
Report Number9681834-2018-00134
Device Sequence Number16355779
Product Code DYB
UDI-Device Identifier04987350717986
UDI-Public04987350717986
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K142183
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberRF*ZM7410GA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2018
Initial Date Received by Manufacturer 06/29/2018
Supplement Date Received by Manufacturer08/02/2018
08/08/2018
Initial Report FDA Received Date07/23/2018
Supplement Report FDA Received Date08/07/2018
08/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient Weight83
-
-