COVIDIEN MFG DC BOULDER COOL-TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number ACT1530
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, the electrode was connected to the generator and the pump was started and ablation was performed.The temperature was unstable and an error occurred after few seconds.The electrode was replaced with new one and the new one worked fine.There was no patient harm.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: evaluation summary: one device was received for evaluation.Visual inspection found no defects.The reported condition was not confirmed.The investigation found the device to function normally and within specifications.The water circulated normally through the sample.The sample read the temperature properly.This unit does not meet the requirements for a manufacturing failure.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOL-TIP
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key7710764
Report Number1717344-2018-00954
Device Sequence Number1107799
Product Code GEI
UDI-Device Identifier10884524000947
UDI-Public10884524000947
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K973297
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberACT1530
Device Catalogue NumberACT1530
Device Lot Number73450195
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2018
Initial Date Received by Manufacturer 07/06/2018
Supplement Date Received by Manufacturer08/08/2018
09/28/2018
Initial Report FDA Received Date07/23/2018
Supplement Report FDA Received Date10/02/2018
10/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-