| Model Number |
AUTOMATE 2550 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
Crushing Injury (1797)
|
| Date of Event |
07/02/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A beckman coulter field service engineer (fse) went onsite on (b)(4) 2018 and evaluated the instrument.The fse did not find any malfunction of the device.The operator failed to depressurize the system before attempting to remove the blockage.The instructions for use clearly state to depressurize the compressor in case of a tube jam.(b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a user incorrectly tried to remove tubes that were blocked/jammed from inside the automate 2550 lab instrument.The user's thumb became "blocked" by the roller resulting in a bruised thumb.The user received a medical review and prescription anti-inflammatory treatment.
|
| |
|