| Model Number |
M0068403940 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/03/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exact patient age is unknown; however the patient was reported to be over 18 years old.(b)(4).The device has been received for analysis; however the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that a flexiva 1000 laser fiber was used during a cystoscopy litholapaxy procedure in the urethra performed on (b)(6) 2018.Reportedly, the laser unit used was 100 watt holmium with laser settings of 1.0 joules and a rate 15.According to the complainant, during the procedure, the tip of the laser fiber broke off inside the patient¿s urethra.The broken fragment was retrieved successfully.The procedure was completed with another flexiva 1000 laser fiber.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be "patient is unaffected by event".
|
| |
|
Additional Manufacturer Narrative
|
|
One flexiva 1000 laser fiber was received for evaluation.Visual examination revealed that the exposed glass fiber tip measured approximately 0.5 mm and appeared used.Additional examination under magnification revealed that the pattern on the tip face was consistent with a fracture indicating that the tip or portion of the tip detached.There was no damage along the body of the fiber.The condition of the returned device confirms the event.The noted damages indicate difficulty was experienced during the procedure and was likely due to anatomical or procedural factors such as tortuous anatomy or maneuvering of the device.Therefore, a review and analysis of all available information indicated that the most probable root cause is operational context.A review of the device history record (dhr) was performed; no anomalies were noted.A search of the complaint database revealed that no similar complaints exist for the specified lot.A labeling review was performed, and from the information available this device was used per the directions for use (dfu)/product label.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that a flexiva 1000 laser fiber was used during a cystoscopy litholapaxy procedure in the urethra performed on (b)(6) 2018.Reportedly, the laser unit used was 100 watt holmium with laser settings of 1.0 joules and a rate 15.According to the complainant, during the procedure, the tip of the laser fiber broke off inside the patient¿s urethra.The broken fragment was retrieved successfully.The procedure was completed with another flexiva 1000 laser fiber.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be "patient is unaffected by event.".
|
| |
|