|
STRYKER TRAUMA KIEL RECONSTRUCTION NAIL R2.0, TI, LEFT T2 RECON Ø11X360 MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Catalog Number |
18461136S |
| Health Effect - Clinical Code |
Irritation (1941)
|
| Date of Event |
01/01/2014
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.Device not available to stryker.
|
| |
|
Event or Problem Description
|
|
The manufacturer became aware of a post market clinical follow-up (pmcf) from (b)(6) hospital (b)(6).The title of this report is ¿analysis of the t2 recon nail used with the antegrade locking mode¿ which was released in june 2018 and is associated with the t2 recon nailing system.Within that report, postoperative complications/ adverse events were reported, which occurred between january 2014 and november 2017.Based on the available device and patient information and a review of complaint handling database it was revealed that the events have not been reported by the hospital or by the author of the report, therefore 15 complaints were initiated retrospectively for different adverse events mentioned in the journal.This product inquiry addresses hardware irritation.One (case id 12) out of 2 cases.The pmcfr states: ¿in two cases (case12 and 26) the hardware was irritating the soft tissue and the removal of the implants was necessary to solve the problem.In the first of these two cases (case 12) the nail sticks out the tip of the greater trochanter in a femoral 32-a fractures.The nail used was 11 mm in diameter and 360 mm in length (18461136s batch code k112936) in a (b)(6) years old male ((b)(6)).The problem was solved by removal of the hardware.¿.
|
| |
|
Search Alerts/Recalls
|
|
|