STRYKER TRAUMA KIEL RECONSTRUCTION NAIL R2.0, TI, LEFT T2 RECON Ø11X360 MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 18461136S
Health Effect - Clinical Code Irritation (1941)
Date of Event 01/01/2014
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.Device not available to stryker.
 
Event or Problem Description
The manufacturer became aware of a post market clinical follow-up (pmcf) from (b)(6) hospital (b)(6).The title of this report is ¿analysis of the t2 recon nail used with the antegrade locking mode¿ which was released in june 2018 and is associated with the t2 recon nailing system.Within that report, postoperative complications/ adverse events were reported, which occurred between january 2014 and november 2017.Based on the available device and patient information and a review of complaint handling database it was revealed that the events have not been reported by the hospital or by the author of the report, therefore 15 complaints were initiated retrospectively for different adverse events mentioned in the journal.This product inquiry addresses hardware irritation.One (case id 12) out of 2 cases.The pmcfr states: ¿in two cases (case12 and 26) the hardware was irritating the soft tissue and the removal of the implants was necessary to solve the problem.In the first of these two cases (case 12) the nail sticks out the tip of the greater trochanter in a femoral 32-a fractures.The nail used was 11 mm in diameter and 360 mm in length (18461136s batch code k112936) in a (b)(6) years old male ((b)(6)).The problem was solved by removal of the hardware.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECONSTRUCTION NAIL R2.0, TI, LEFT T2 RECON Ø11X360 MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
MDR Report Key7711356
Report Number0009610622-2018-00381
Device Sequence Number11017645
Product Code HSB
UDI-Device Identifier07613153316633
UDI-Public07613153316633
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K032898
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number18461136S
Device Lot NumberK098BB7 / K112936
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2018
Initial Report FDA Received Date07/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age29 YR
Patient Weight80
-
-