BOSTON SCIENTIFIC - GALWAY WALLFLEX¿ BILIARY RX STENT SYSTEM; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00570730
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has been received for analysis; however the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that a wallflex biliary rx partially covered stent was to be used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on (b)(6) 2018.Reportedly, the patient¿s anatomy was not tortuous.According to the complainant, during the procedure, there was difficulty exiting the guidewire at the guidewire rx port.After several attempts, the guidewire was able to finally pass through the rx port.During stent deployment, the inner catheter broke in half inside the patient, and the stent was not deployed.The wallflex biliary device was removed from the patient and the procedure was not completed.There were no patient complications reported as a result of this event.The patient's condition was reported to be fine.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that a wallflex biliary rx partially covered stent was to be used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on (b)(6) 2018.Reportedly, the patient¿s anatomy was not tortuous.According to the complainant, during the procedure, there was difficulty exiting the guidewire at the guidewire rx port.After several attempts, the guidewire was able to finally pass through the rx port.During stent deployment, the inner catheter broke in half inside the patient, and the stent was not deployed.The wallflex biliary device was removed from the patient and the procedure was not completed.There were no patient complications reported as a result of this event.The patient's condition was reported to be fine.
 
Additional Manufacturer Narrative
A wallflex biliary delivery system was received for analysis; the stent was not returned.The delivery system was received in two pieces.Visual examination of the returned device found the clear outer sheath at the distal end was slightly kinked.The inner member was found detached, kinked and damaged.No other issues with the delivery system were noted.The damages noted with the returned device were consistent with excessive force being applied to the delivery system.The investigation concluded that the observed failures were likely due to anatomical or procedural factors encountered during the procedure which limited the performance of the device.Therefore, the most probable root cause is operational context.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.A search of the complaint database confirmed that no similar complaints exist for the specified lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX¿ BILIARY RX STENT SYSTEM
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7711704
Report Number3005099803-2018-02289
Device Sequence Number2580775
Product Code FGE
UDI-Device Identifier08714729759324
UDI-Public08714729759324
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K140630
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/03/2020
Device Model NumberM00570730
Device Catalogue Number7073
Device Lot Number0021949235
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2018
Initial Date Received by Manufacturer 06/29/2018
Supplement Date Received by Manufacturer08/07/2018
Initial Report FDA Received Date07/23/2018
Supplement Report FDA Received Date08/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-