| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
06/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-14- insufficient blood sample applied to strip (user) test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure the customer's condition since the initial call and that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for e-2 error: sample not detected or using wrong test strip.The e-2 error occurred after the blood sample was applied.At the time of the call the customer stated she had distorted vision; medical attention was not needed at the time of the call.Customer is using correct test strips and product is stored according to the specification.During the call on (b)(6) 2018 a blood test was performed by the customer and produced e-2 error using truemetrix meter.The test strip lot manufacturer's expiration date is 09/23/2019 and open vial date is (b)(6) 2018.
|
| |
|