WRIGHTS LANE SYNTHES USA PRODUCTS LLC 10MM/130 DEG TI CANN TFNA 380MM/RIGHT - STERILE; ROD, FIXATION, INTRAMEDULLARY
|
Back to Search Results |
|
| Model Number |
04.037.058S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.A review of the device history records has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that patient complained of hip pain on an unknown date and it was discovered the nail was broken.The trochanteric femoral nail advanced (tfna) nail was originally implanted on (b)(6) 2018.On (b)(6) 2018, the revision surgery was done due to a broken (tfna ) nail, which was reported as broken through the proximal helical blade hole.It was noted the broken nail did not generate any fragments.The nail was removed and revised with a larger (tfna) nail.The procedure was successfully completed with no surgical delay.Concomitant device reported: unknown distal locking screw (part unknown, lot unknown, quantity 1); helical blade (part 04.038.310, lot h512772, quantity unknown).This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device history record (dhr) review: item: 04.037.058s; lot: h252968; mfg qty: 4; mfg date: 13 dec 2016; mfg location: monument; scn: 13260; exp date: 30 nov 2026; product was inspected and passed per inspection sheet.Raw material lot h123944 was used.Item: 21127; lot: h123944; qty rec'd: 2941 pounds; rec'd date: 07 jun 2016; rec'd from: metalwerks pmd, incorporated; certificate of analysis received from metalwerks and raw material receiving/putaway checklist indicate product meets specifications.There were no nonconformances (ncs) or reworks generated during the processing of this either the product or the material lots.No nonconformance records (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Device manufacture date updated as per the dhr review.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|