| Model Number |
H7493926012250 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.(b)(4).
|
| |
|
Event or Problem Description
|
|
Same case as mdr id: 2134265-2018-06647.It was reported that stent damage occurred.The target lesion was located in the calcified mid right coronary artery.A 2.50 x 16mm synergy ii drug-eluting stent was advanced but failed to cross the lesion and it was noted that the stent was damaged.The device was removed and a 2.50 x 12mm synergy ii drug-eluting stent was advanced but still failed to cross the lesion and damaged was also noted.The device was removed from the patient.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: synergy ii us mr 2.50 x 12mm stent delivery system was returned for analysis.A visual examination of the crimped stent identified damage.Proximal strut 4 was lifted and bent in a proximal direction.The remainder of the stent was undamaged.The undamaged crimped stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the device revealed multiple hypotube kinks.An examination of the shaft polymer extrusion found no issues.The bi-component bond showed no signs of damage.The tip was visually and microscopically examined and damage was noted.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
| |
|
Event or Problem Description
|
|
Same case as mdr id: 2134265-2018-06647.It was reported that stent damage occurred.The target lesion was located in the calcified mid right coronary artery.A 2.50 x 16mm synergy ii drug-eluting stent was advanced but failed to cross the lesion and it was noted that the stent was damaged.The device was removed and a 2.50 x 12mm synergy ii drug-eluting stent was advanced but still failed to cross the lesion and damaged was also noted.The device was removed from the patient.No patient complications were reported.
|
| |
|