BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number H7493926012250
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.(b)(4).
 
Event or Problem Description
Same case as mdr id: 2134265-2018-06647.It was reported that stent damage occurred.The target lesion was located in the calcified mid right coronary artery.A 2.50 x 16mm synergy ii drug-eluting stent was advanced but failed to cross the lesion and it was noted that the stent was damaged.The device was removed and a 2.50 x 12mm synergy ii drug-eluting stent was advanced but still failed to cross the lesion and damaged was also noted.The device was removed from the patient.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: synergy ii us mr 2.50 x 12mm stent delivery system was returned for analysis.A visual examination of the crimped stent identified damage.Proximal strut 4 was lifted and bent in a proximal direction.The remainder of the stent was undamaged.The undamaged crimped stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the device revealed multiple hypotube kinks.An examination of the shaft polymer extrusion found no issues.The bi-component bond showed no signs of damage.The tip was visually and microscopically examined and damage was noted.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
Same case as mdr id: 2134265-2018-06647.It was reported that stent damage occurred.The target lesion was located in the calcified mid right coronary artery.A 2.50 x 16mm synergy ii drug-eluting stent was advanced but failed to cross the lesion and it was noted that the stent was damaged.The device was removed and a 2.50 x 12mm synergy ii drug-eluting stent was advanced but still failed to cross the lesion and damaged was also noted.The device was removed from the patient.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7711884
Report Number2134265-2018-06642
Device Sequence Number2644279
Product Code NIQ
UDI-Device Identifier08714729840039
UDI-Public08714729840039
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/13/2019
Device Model NumberH7493926012250
Device Catalogue Number39260-1225
Device Lot Number0021763446
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2018
Initial Date Received by Manufacturer 07/03/2018
Supplement Date Received by Manufacturer08/09/2018
Initial Report FDA Received Date07/23/2018
Supplement Report FDA Received Date08/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-