MEDTRONIC NAVIGATION, INC. (LOUISVILLE) MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM; EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION COMPACT
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/29/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.Device manufacturing date is unavailable.No investigation was performed as the issue was resolved during initial troubleshooting.
 
Event Description
Medtronic received information regarding a navigation system being used during a functional endoscopic sinus surgery (fess).It was reported that the axiem box was not communicating with the system.Rebooting the system multiple times resolved the issue.This occurred intra/peri-operatively with less than one hour delay to surgical time.There was no known impact to patient outcome.
 
Manufacturer Narrative
Device manufacturing date updated to proper value.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM
Type of Device
EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
heather davis
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7711962
MDR Text Key114802296
Report Number1723170-2018-03655
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00643169672956
UDI-Public00643169672956
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFUSION COMPACT
Device Catalogue Number9735602
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/29/2018
Initial Date FDA Received07/23/2018
Supplement Dates Manufacturer Received08/03/2018
Supplement Dates FDA Received08/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/22/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-