MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problem No Display/Image (1183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/22/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site. No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding a navigation device being used for a functional endoscopic sinus surgery (fess) procedure. The issue occurred intraoperatively while setting up equipment and delayed the surgery by less than one hour. It was reported that the monitor would not show imaged, it was just a black screen. The patient scans had been loaded onto the system in the morning with no difficulties, but when the system was brought into the room, the issue occurred. The system was rebooted, and the issue was resolved. The surgery was completed with navigation and there was no known impact on patient outcome. Medtronic representative reported that the site biomedical engineer opened the computer and found that the graphics card was failing.
 
Manufacturer Narrative
Additional information: site found that the 3v li-ion coin cell battery on the mother board was nearly dead. The site replaced the battery (which pops in and out of its cradle) and the problem went away. The site tested it several times to ensure the blanking didn¿t come back. If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFUSION NAVIGATION SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7712385
MDR Text Key114802588
Report Number1723170-2018-03659
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/06/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberFUSION
Device Catalogue Number9733560XOMR
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/03/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/22/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-