| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/12/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional manufacturer narrative/corrected data: conclusion code - complications associated with the use of the gore® tag® thoracic endoprosthesis may include but are not limited to endoleak.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2015, this patient underwent endovascular repair of a descending thoracic aortic aneurysm using conformable gore® tag® thoracic endoprostheses.On (b)(6) 2017, a type ib endoleak was identified.It was reported that the endoleak was caused by patient anatomy and distal end of the most distal device not having good aortic seal.On (b)(6) 2018, a reintervention occurred whereby a two gore® excluder® aaa endoprosthesis aortic extender components were implanted.At the conclusion of the intervention, an aortogram was performed and a very slow endoleak was visualized.The procedure was concluded with no further action.
|
| |
|