Catalog Number 71453201
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/27/2018
Type of Reportable Event Malfunction
Event or Problem Description
During tka procedure, the insert could not be locked into baseplate (71420162/17lm00130).Backup (10mm) insert was implanted as replacement.No patient injury or delay was reported.
 
Additional Manufacturer Narrative
The associated legion ps high flexion insert was returned and evaluated.A lab analysis noted that no major damage is seen on the insert.The locking detail is slightly deformed, likely due to removal.Minor scratches on the articulating surface of the insert.There were no observations of material or manufacturing deviations in the course of this investigation.A dimensional inspection was attempted, but the deformation of the locking mechanism would not allow for accurate measurement.Our investigation including a review of complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.A review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN PS HIGH FLEX XLPE SZ 1-2 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7712721
Report Number1020279-2018-01357
Device Sequence Number1370264
Product Code JWH
UDI-Device Identifier00885556033784
UDI-Public00885556033784
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71453201
Device Lot Number17LT95771
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/27/2018
Supplement Date Received by Manufacturer06/27/2018
Initial Report FDA Received Date07/23/2018
Supplement Report FDA Received Date12/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-