| Catalog Number |
0350800001 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Mfg date: the device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device began smoking during the case.
|
| |
|
Additional Manufacturer Narrative
|
|
The device manufacture date is not known.Alleged failure: started smoking during case.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause could be main board non conforming components.The product was returned for investigation and the failure mode was confirmed and will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the device began smoking during the case.
|
| |
|