| Catalog Number |
EUP3015X |
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A euphora balloon catheter was intended to be used to treat a lesion.The device was inspected before use with no issues noted.Negative prep/purging was performed on the device prior to use with no issues noted.It is reported that during the inflation of a euphora balloon catheter, the doctor realized that the balloon inflated partially, but was not inflating fully.The physician thought there was a hole, so they removed it from the patient's body.When it was removed he inflated it and confirmed that there was a hole.It is not known if the hole was present prior to use, from prepping or from insertion.The procedure was completed with a new 3.0x15mm euphora.No patient injury is reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|