Complaint confirmed.The received screw was undamaged and did not show signs of biodegradation.Absorption of biodegradable implants begins at implantation as a result of normal hydrolysis of the implant.The rate at which the implant degrades and is reabsorbed can be affected by numerous factors including, but not limited to implantation site vascularity, stress on the implants, surgical techniques employed during implantation, patient factors (such as age, immune system status, bone quality, body chemistry and sensitivity to implant materials), and patient post-op compliance to rehabilitation protocols.There is no minimum or maximum established resorption rates for bioabsorbable implants.Bioabsorbable implants are designed with material properties required to maintain necessary properties throughout the healing process under normal patient conditions.
|