Catalog Number UNK HIP ACETABULAR CUP |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/12/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Depuy 18-53 detachment cup-insert.Revision surgery performed on (b)(6).
|
|
Event Description
|
It was reported that images show a poly liner with some of the anti rotational lugs sheared off.It also shows excessive wear of the liner to such an extent that the poly liner has a fracture in its side.The x ray image shows the head over to one side of the shell, implying that the liner has disassociated.
|
|
Manufacturer Narrative
|
Product complaint #: (b)(4).Investigation summary: no device was returned.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate. .
|
|