• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US AVIATOR VARIABLE SELF DRILLING SCREW 4X14MM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US AVIATOR VARIABLE SELF DRILLING SCREW 4X14MM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY Back to Search Results
Catalog Number 48824014
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Failure of Implant (1924); Injury (2348); No Code Available (3191)
Date of Event 07/02/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent revision surgery to replace a plate and screws, as the screws were backing out.
 
Additional Manufacturer Narrative
The customer reported event was confirmed by the provided x-ray.A revision surgery occurred and the implants were returned for examination.The visual inspection found the screw shave no anomalies associated.Manufacturing files were reviewed and no anomalies were found.The exact root cause cannot be determined.The locking mechanism of the screw is on the plate.The potential root causes include, poor surgical candidate or bone quality, excessive post-op activity by patient (not recommended by surgeon), patient does not follow surgeon instructions, patient over exerts.
 
Event or Problem Description
It was reported that the patient underwent revision surgery to replace a plate and screws, as the screws were backing out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVIATOR VARIABLE SELF DRILLING SCREW 4X14MM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key7714231
Report Number0009617544-2018-00179
Device Sequence Number1107808
Product Code KWQ
UDI-Device Identifier07613252152606
UDI-Public07613252152606
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number48824014
Device Lot NumberWE9
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2018
Initial Date Received by Manufacturer 07/02/2018
Supplement Date Received by Manufacturer09/28/2018
Initial Report FDA Received Date07/24/2018
Supplement Report FDA Received Date10/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-