• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FUJIFILM CORPORATION FUJIFILM; GASTROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FUJIFILM CORPORATION FUJIFILM; GASTROSCOPE Back to Search Results
Model Number EG-600WR V2
Medical Device Problem Code Component or Accessory Incompatibility (2897)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A fujifilm field service engineer on site found no evidence that the brush part was still in the scope.The scope was sent to a fujifilm repair center for inspection -- there were no defects/malfunctions observed during the inspection.Although the cause of the incident is tied to the use of an un-approved brush, this incident is being considered reportable by fujifilm in an abundance of caution.
 
Event or Problem Description
A medline brush (not validated or recommended for use with fujifilm scopes) was used to clean a fujifilm gastroscope model eg-600wr v2.Part of this brush broke during the cleaning process.The customer is not certain if the broken brush part had gotten stuck in the scope or was flushed out of the scope.No evidence of the broken brush part was found after subsequent reprocessing cycles.The scope was used on other patients, with no reports of infection.About a week later, the customer informed fujiflim that the brush had been found; however, the customer refused to send in the brush.Fujifilm was informed that the specific medline brush model was currently being recalled by the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUJIFILM
Common Device Name
GASTROSCOPE
Manufacturer (Section D)
FUJIFILM CORPORATION
798 miyanodai kaisei-machi
ashigarakami-gun, kanagawa 258-8 538
JA  258-8538
MDR Report Key7714294
Report Number3001722928-2018-00056
Device Sequence Number16410440
Product Code FDS
UDI-Device Identifier04547410337402
UDI-Public(01)04547410337402(11)170201
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K132210
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEG-600WR V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/29/2018
Initial Report FDA Received Date07/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-