| Catalog Number |
6001-088-427 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during testing at the user facility it was observed that the tip of the device was broken off.No medical intervention and no adverse consequences were reported with this event.As this event occurred during testing at the user facility, there was no patient involvement and no delay to a surgical procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: follow-up report submitted to document device evaluation results.
|
| |
|
Event or Problem Description
|
|
It was reported that during testing at the user facility it was observed that the tip of the device was broken off.No medical intervention and no adverse consequences were reported with this event.As this event occurred during testing at the user facility, there was no patient involvement and no delay to a surgical procedure.
|
| |
|