WRIGHTS LANE SYNTHES USA PRODUCTS LLC MATRIXMANDIBLE 1.5MM DRILL BIT J-LATCH FOR 03.503.045/.047; DRILL, BONE, POWERED Back to Search Results
Model Number 03.503.476
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 05/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient information not available for reporting.Device is an instrument and is not implanted/explanted.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported patient underwent an unknown oral procedure on (b)(6) 2018.During the procedure, the surgeon used the matrixmandible 1.5mm drill bit intra-orally through a 2.0mm cannula.The mouth area was irrigated but the cannula was not.The drill speed was reported as 28,000 rpm.Surgeon stated the drill bit became very hot and suspects the fit between the drill bit and cannula may be too tight, which may have caused friction and heat.While it was initially reported that patient was not injured, surgeon later reported the patient¿s skin surrounding the drill and trocar was hypertrophic (burned) associated with this event.Surgeon stated patient may require plastic surgery consults and/or treatment of the hypertrophic area.Concomitant devices reported: 2.0mm cannula & obturator (397.213, lot number unknown, quantity 1).This report is for one (1) matrixmandible 1.5mm drill bit.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MATRIXMANDIBLE 1.5MM DRILL BIT J-LATCH FOR 03.503.045/.047
Common Device Name
DRILL, BONE, POWERED
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7714614
Report Number2939274-2018-53046
Device Sequence Number1107739
Product Code DZI
UDI-Device Identifier10887587012971
UDI-Public(01)10887587012971
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K043310
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.503.476
Device Catalogue Number03.503.476
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/18/2018
Supplement Date Received by Manufacturer11/02/2020
Initial Report FDA Received Date07/24/2018
Supplement Report FDA Received Date11/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-