| Model Number |
METRIX AIR |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.(b)(4).Manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 199, 275, 230 and 343 mg/dl.The customer's expected fasting blood glucose test result range is 166 - 170 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 05/29/2019 and open vial date is more than three months.The meter memory was reviewed for previous test result history: result 1 :199mg/dl date: (b)(6) 2018 time:8:49 am fasting; result 2 :275mg/dl date: (b)(6) 2018 time:8:04am fasting; result 3 :230mg/dl date: (b)(6) 2018 time:8:03am fasting; result 4 :343mg/dl date: (b)(6) 2018 time:12:18pm fasting.
|
| |
|
Search Alerts/Recalls
|