• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO TPS UNIDIRECTIONAL FOOTSWITCH; UNIT, OPERATIVE DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO TPS UNIDIRECTIONAL FOOTSWITCH; UNIT, OPERATIVE DENTAL Back to Search Results
Catalog Number 5100007000
Medical Device Problem Code Device Remains Activated (1525)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Reported events: one event was reported for this quarter.Product return status one device was received for evaluation.Evaluation status: one event was not confirmed during testing; the device was found to be within specifications.Product disposition: this device is not repairable and was not returned to the account.Additional information: there were no remedial actions taken.This device is not labeled for single-use.
 
Event or Problem Description
This report summarizes 1 malfunction event in which the device had run-on.One event had patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPS UNIDIRECTIONAL FOOTSWITCH
Common Device Name
UNIT, OPERATIVE DENTAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key7714807
Report Number0001811755-2018-01204
Device Sequence Number13702153
Product Code EIA
UDI-Device Identifier04546540401724
UDI-Public04546540401724
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number5100007000
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2018
Initial Report FDA Received Date07/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-