| Catalog Number |
TGU404010 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Aneurysm (1708)
|
| Date of Event |
10/20/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: tgu404010 udi: (b)(4), tgu404010 udi: (b)(4), tgu404010 udi: (b)(4).It is not known which of the tgu404010 devices were involved in the event, therefore all were investigated.The viabahn device is being reported separately.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2016, the patient underwent treatment of a thoracoabdominal aortic aneurysm whereby conformable gore® tag® thoracic endoprostheses and gore® viabahn® endoprostheses with heparin bioactive surface were implanted.On (b)(6) 2018 the patient exhibited a gutter leak around the left renal artery.On (b)(6) 2018 the patient's left renal artery was stented and a left renal artery branch in the gutter space was embolized.The patient tolerated the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event.Updated event.Updated conclusion code 1: 22: according to the conformable gore tag® thoracic endoprosthesis instructions for use, adverse events that may occur include, but are not limited to endoleak, aneurysm enlargement and reoperation.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6)2016, the patient underwent treatment of a thoracoabdominal aortic aneurysm whereby conformable gore® tag® thoracic endoprostheses and gore® viabahn® endoprostheses with heparin bioactive surface were implanted.On (b)(6)2017 the maximum aortic diameter was 67mm.On (b)(6)2017 the maximum aortic diameter was 75mm.On (b)(6)2018 the patient exhibited a gutter leak around the left renal artery.On (b)(6)2018 the patient's left renal artery was stented and a left renal artery branch in the gutter space was embolized.The patient tolerated the procedure.On (b)(6)2018 the patient was recovering from his surgery with severe deconditioning.
|
| |
|
Search Alerts/Recalls
|