• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO STRYKER PRECISION OSCILLATING TIP SAW; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO STRYKER PRECISION OSCILLATING TIP SAW; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 6209000000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Reported events: 1 event was reported for this quarter.Product return status: 1 device was available for evaluation but has not yet been received.Additional information: there were no remedial actions taken on this device.This device is not labeled for single-use.
 
Event or Problem Description
This report summarizes 1 malfunction event in which the device had a broken cutting accessory still attached.One event had patient involvement; no patient impact.
 
Additional Manufacturer Narrative
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.  supplemental rationale corrected data: 1 event was initially reported.1 previously reported event is included in this follow-up record.  product return status 1 device was received.Evaluation status 1 reported event was not confirmed during testing.Additional information 1 device is not labeled for single-use.1 device was not reprocessed and reused.
 
Event or Problem Description
This report summarizes 1 malfunction event in which the device had a broken cutting accessory still attached.1 event had patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRYKER PRECISION OSCILLATING TIP SAW
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key7715080
Report Number0001811755-2018-01325
Device Sequence Number904571
Product Code HWE
UDI-Device Identifier07613154488285
UDI-Public07613154488285
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number6209000000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/02/2018
Supplement Date Received by Manufacturer03/02/2018
Initial Report FDA Received Date07/24/2018
Supplement Report FDA Received Date01/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-