| Catalog Number |
M0035434150 |
| Medical Device Problem Codes |
Migration (4003); Premature Separation (4045)
|
| Health Effect - Clinical Codes |
Thrombus (2101); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
06/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Subject device is not available.
|
| |
|
Event or Problem Description
|
|
It was reported that during stent-assisted coil embolization of a ruptured aneurysm located in the left posterior communicating artery (pcomm), midway through deployment of the second coil (subject device), the delivery wire prematurely detached from the coil and the coil migrated.Angiograms showed thrombus propagation in the left pcomm and blood flow stasis in the distal branches of the left pcomm.Antiplatelet drug (tirofiban) was administered.A stent was partially deployed in the terminal ica to complete the coiling and then the stent was fully deployed to jail the migrated coil.The distal part of the coil however still migrated outside the deployed stent into the to the internal carotid artery (ica) bifurcation.Movement of the coil with blood pulsation was noted.Another stent was deployed to secure and jail the still distally migrated coil.Thrombus in the left pcomm was resolved and blood flow was restored.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record could not be performed because the lot number was not reported.The subject device is not available.In the case of this complaint it is most probable that procedural or anatomical factors encountered during the clinical procedure contributed to the reported premature detachment, possibly leading the main coil migration, and patient thrombosis (treated vessel).This however cannot be conclusively determined.Therefore, an assignable cause of 'undeterminable' will be assigned to these reported issues as the device was not returned and review and analysis of all available information fails to indicate an assignable cause or probable assignable cause.Vessel thrombosis is a known risk associated with endovascular procedures and is noted as such in the device directions for use (dfu).Therefore, an assignable cause of anticipated procedural complications has been assigned to this patient thrombosis.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during stent-assisted coil embolization of a ruptured aneurysm located in the left posterior communicating artery (pcomm), midway through deployment of the second coil (subject device), the delivery wire prematurely detached from the coil and the coil migrated.Angiograms showed thrombus propagation in the left pcomm and blood flow stasis in the distal branches of the left pcomm.Antiplatelet drug (tirofiban) was administered.A stent was partially deployed in the terminal ica to complete the coiling and then the stent was fully deployed to jail the migrated coil.The distal part of the coil however still migrated outside the deployed stent into the to the internal carotid artery (ica) bifurcation.Movement of the coil with blood pulsation was noted.Another stent was deployed to secure and jail the still distally migrated coil.Thrombus in the left pcomm was resolved and blood flow was restored.
|
| |
|