| Catalog Number |
5400110000 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Reported events: 4 events were reported for this quarter.Product return status: 4 devices were received for evaluation.Evaluation status: 1 event was confirmed during testing, - 1 device was found to be affected by friction problem (rotor rub), 3 device evaluations are in progress.Product disposition: the device was repaired and returned to the user facility.Additional information: there were no remedial actions taken on this device.This device is not labeled for single-use.
|
| |
|
Event or Problem Description
|
|
This report summarizes 4 malfunction events in which the device reportedly overheated; 3 events had no patient involvement; no patient impact; 1 event had no known patient involvement or patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Supplemental rationale corrected data: 4 total events were initially reported.3 previously reported events are included in this follow-up record. product return status 3 devices were received. evaluation status confirmed 3 reported events were confirmed during testing.- 1 device was found to be affected by rotor rub/damage.- 1 device was found to be affected by damaged rotor assembly.- 1 device was found to be affected by worn rotor stop. additional information 3 devices were not labeled for single-use.3 devices were not reprocessed and reused.
|
| |
|
Event or Problem Description
|
|
This report summarizes 4 malfunction events in which the device reportedly overheated.3 events had no patient involvement; no patient impact.1 event had no known patient involvement or patient impact.
|
| |
|
Search Alerts/Recalls
|