| Model Number |
TRUETRACK |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage(kitchen).(b)(4).Manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose results.Wife is calling on behalf of the customer.The customer is concerned with results obtained of 227, 382, 300 and 489mg/dl.The expected fasting blood glucose test result range is 90 to 120mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored by customer according to specification in the bathroom.Test strip lot manufacturer's expiration date is 07/23/2020 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly) result 1:227mg/dl date: (b)(6) 2017 time:12:04a n/fasting; result 2:382mg/dl date: (b)(6) 2017 time:12:00a n/ fasting; result 3:300mg/dl date: (b)(6) 2017 time:7:05a n/fasting; result 4:489mg/dl date: (b)(6) 2017 time:6:18p n/fasting; result 5:166mg/dl date: (b)(6) 2017 time:6:41p n/fasting; strips were exposed to extreme heat for 3 hours on (b)(6) 2018.
|
| |
|
Search Alerts/Recalls
|