Additional Manufacturer Narrative
|
|
This mdr report is part of the malfunction summary reporting program, exemption number e2015055. reported events: two events were reported for this quarter.Product return status: two devices were received for evaluation.Evaluation status: two device evaluations are in progress.Additional information: there were no remedial actions taken.This device is not labeled for single-use.
|
| |
|
Event or Problem Description
|
|
This report summarizes 2 malfunction events in which the device was reportedly leaking.Two events had no patient involvement; no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Supplemental rationale corrected data: 2 total events were initially reported.2 previously reported events are included in this follow-up record. product return status 2 devices were received. evaluation status not confirmed 2 reported events were not confirmed during testing; however: - 2 devices were found to be affected by a split weld. additional information 2 devices were not labeled for single-use.2 devices were not reprocessed and reused.
|
| |
|
Event or Problem Description
|
|
This report summarizes 2 malfunction events in which the device was reportedly leaking.2 events had no patient involvement; no patient impact.
|
| |
|