Catalog Number 6206000000
Medical Device Problem Code Device Remains Activated (1525)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Reported events : 1 event was reported for this quarter.Product return status: 1 device was received for evaluation.Evaluation status: 1 event was confirmed during testing, - 1 device was found to be affected by an electrical problem.Product disposition: the device was repaired and returned to the user facility.Additional information: there were no remedial actions taken on this device.This device is not labeled for single-use.
 
Event or Problem Description
This report summarizes 1 malfunction event in which the device had run-on; 1 event had no patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYSTEM 6 RECIP
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key7715555
Report Number0001811755-2018-01512
Device Sequence Number996325
Product Code HWE
UDI-Device Identifier07613154392360
UDI-Public7613154392360
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number6206000000
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2018
Initial Report FDA Received Date07/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-