| Catalog Number |
6203000000 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr report is part of the malfunction summary reporting program, exemption number e2015055. reported events: 9 events were reported for this quarter.Product return status : 9 devices were received for evaluation.Evaluation status : 8 events were confirmed during testing, - 7 devices were found to be affected by detached/missing components, - 1 device was found to be affected by 3rd party repairs, 1 device evaluation is in progress.Product disposition : the devices were repaired and returned to the user facility. additional information : there were no remedial actions taken on this device.This device is not labeled for single-use.
|
| |
|
Event or Problem Description
|
|
This report summarizes 9 malfunction events in which the device had a component detach; 8 events had no patient involvement; no patient impact; 1 event had patient involvement; no patient impact.
|
| |
|
Event or Problem Description
|
|
This report summarizes 9 malfunction events in which the device had a component detach.8 events had no patient involvement; no patient impact.1 event had patient involvement; no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Supplemental rationale corrected data: 9 total events were initially reported.1 previously reported event is included in this follow-up record. product return status - 1 device was received. evaluation status confirmed 1 reported event was confirmed during testing.- 1 device was found to be affected by trigger assembly sheared at the cap screws and wear on the release collar. additional information 9 devices were not labeled for single-use.9 devices were not reprocessed and reused.
|
| |
|