| Model Number |
D334DRM |
| Medical Device Problem Code |
Failure to Interrogate (1332)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
| Date of Event |
05/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 1882tc, implanted: (b)(6) 2013.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) was explanted and replaced due to recommended replacement time (rrt).The icd subsequently tested out of specification and was unable to be interrogated.In addition, it was reported that the superior vena cava (svc) lead impedance warning was triggered for the right ventricular (rv) lead.The rv lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and analyzed.Analysis of the device revealed normal battery depletion.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|