| Model Number |
4A5044 |
| Medical Device Problem Code |
Inaccurate Flow Rate (1249)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter would not drain after insertion.The catheter was replaced and no medical intervention was required.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.Evaluation of the sample noted no obvious defects.The catheter was placed in a clear plastic bag and a slip tip syringe was connected to the drainage funnel and used to flush the catheter.10cc of fluid was able to pass through the catheter and no particulate was found in the plastic bag.The tip and eye of the device was measured and compared to the recommended.The eye and tip length were found to be within tolerances.The french size was measured to be 14 fr.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use was found adequate and states the following: ¿bard® touchless¿ plus unisex pre-lubricated urethral catheter kit contents: rubber* urethral catheter 1100cc collection bag 3 povidone-iodine swabs 2 ambidextrous gloves waterproof underpad instructions for use: preparing for catheterization: 1.Open kit and remove contents.2.The gloves are not necessary to maintain aseptic conditions during the procedure.The gloves are for the operator's protection.Male/female catheterization: 1.Remove cap from insertion tip while squeezing tabs.Save cap for closing the bag.2.Slide the catheter halfway into the insertion tip.3.Male - prepare the male urethral meatus and the surrounding area with povidone-iodine swabs provided.Female - prepare the female urethral meatus by holding the outer labia apart and prep the urethral meatus and surrounding area with povidone iodine swabs provided.4.Male - holding the penis, advance the insertion tip into the urethra no further than the flange base.Female - spread the inner labia, advance the insertion tip into the urethra no further than the flange base.Release the inner labia.5.Place your non-dominant hand on the finger control guide to stabilize catheter in urethra.With your dominant hand, grasp catheter through bag approximately 1" below the finger control guide and push catheter into urethra.The catheter should be introduced by short, repetitive pushing motions.Repeat motions until catheter reaches bladder and urine starts to flow.6.Allow urine to flow freely, making certain the catheter guide is elevated at least 4" above lower portion of the bag.Allow flow until bladder is empty or until bag is filled.Precaution: it is recommended that the collection bag be held.The catheter could possibly separate from the collection bag when urine increases weight of the bag.7.Withdraw the catheter from the urethra.Remove the remaining portion of the catheter from the collection bag by pulling it through the insertion tip.Note: the catheter is designed to pass through the insertion tip.8.The filled collection bag may be closed by replacing the cap over the insertion tip.Make sure the cap snaps on securely.Specimen collection: to collect a specimen, obtain a sample cup/tube and alcohol wipe.Remove guide tip by pulling tab(s) upward from bag.Wipe port with alcohol.Pour specimen through port at top of bag into cup/tube.Draining bag: remove insertion tip by pulling tab(s) upward from bag.Drain urine through port at top of bag.".
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter would not drain after insertion.The catheter was replaced and no medical intervention was required.
|
| |
|