• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 10MM / TI CANN FRN / GT 400MM / RIGHT ¿ STERILE; ROD,FIXATION,INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 10MM / TI CANN FRN / GT 400MM / RIGHT ¿ STERILE; ROD,FIXATION,INTRAMEDULLARY Back to Search Results
Model Number 04.033.070S
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 06/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device was not explanted.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that patient underwent a procedure to implant a 10mm cannulated femoral recon nail (frn) on (b)(6) 2018.During the procedure, the nail became lodged in the patient¿s proximal femur, as the opening reamer of the nail did not allow for the nail to be fully inserted.Surgeon attempted to back the nail out but was not successful.Surgeon was forced to proceed with insertion of the nail and was eventually able to seat the nail, but only low enough to allow for one (1) proximal femoral recon screw to be implanted.Surgeon was then able to use the nail¿s distal locking options to complete the procedure with a delay of approximately 60 minutes.This report is for one (1) 10mm cannulated frn nail.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: event: additional information provided.Lot number is now known.Part: 04.033.070s; lot: l828180; manufacturing site: bettlach; release to warehouse date: april 06, 2018; expiry date: february 28.2028.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all material, dimensional and visual criteria at the time of release with no issues documented during the manufacturing process.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The surgeon reamed with a 14mm cannulated opening reamer for the proximal diameter of the nail.Then we reamed up to a 12mm for a 10mm nail.
 
Event or Problem Description
Further it was reported that one of the proximal recon screws was unable to be implanted due to the height of the nail in the femur.Here was no adverse effect on the patient reported.Concomitant devices reported: titanium recon screw (part#-04.003.031s, lot#-unknown, quantity 1), titanium recon screw (part#-04.005.534s, lot#-unknown, quantity 1), titanium recon screw (part#-04.005.540s, lot#-unknown, quantity 1).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Expiration date.Added concomitant devices and therapy date.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10MM / TI CANN FRN / GT 400MM / RIGHT ¿ STERILE
Common Device Name
ROD,FIXATION,INTRAMEDULLARY
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7716182
Report Number2939274-2018-53066
Device Sequence Number11042901
Product Code HSB
UDI-Device Identifier10886982272027
UDI-Public(01)10886982272027
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K172157
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.033.070S
Device Catalogue Number04.033.070S
Device Lot NumberL828180
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/28/2018
Supplement Date Received by Manufacturer08/01/2018
08/01/2018
Initial Report FDA Received Date07/24/2018
Supplement Report FDA Received Date08/22/2018
08/24/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age18 YR
Patient Weight60
-
-