| Model Number |
BVX-270B |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Partial thickness (Second Degree) Burn (2694)
|
| Date of Event |
06/08/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported was not attributable to a device malfunction.The doctor alleged that he started to use the returned electrode pad along with the electrosurgical unit and j-plasma handpiece and noted that the patient experienced a burn in the incision site during a subdermal procedure.The engineering team completed an investigation to determine if there are differences between rf current delivered and tissue temperatures achieved with and without a return electrode connected to the patient when a j-plasma handpiece is used in conjunction with the generator.This investigation concluded that the addition of a return electrode to unit had no impact on the amount of rf current being delivered to the patient.Rf current was comparable for all points measured with or without a return electrode connection.No noticeable increase in rf current was observed by the introduction of a grounding pad to the patient during j-plasma procedure vs no grounding pad being used.It is concluded that the addition of the return electrode had no impact on the average and maximum tissue temperature measurements.
|
| |
|
Event or Problem Description
|
|
Complaint alleged that the patient received a burn in the incision sites right and left area of the abdominal after a j-plasma procedure.
|
| |
|
Search Alerts/Recalls
|