| Catalog Number |
4004C0513 |
| Medical Device Problem Codes |
Unraveled Material (1664); Physical Resistance/Sticking (4012); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the pc400 embolization coil was stuck inside the lantern lumen.The proximal constraint sphere was not present on the proximal end of the pc400 embolization coil.Conclusions: evaluation of the returned lantern revealed that the catheter was kinked and ovalized.Further evaluation revealed that the pc400 identified in the complaint was stuck inside the lantern.If the lantern is forcefully advanced through a tortuous anatomy, damage such as a kink and ovalization may occur.The kink and ovalization likely contributed to the resistance experienced while advancing the guidewire and subject pc400 into the lantern.If a coil is advanced through an ovalized catheter, it may become stuck inside the catheter ovalization.If the coil is then forcefully retracted against resistance, it may result in the coil detaching within the catheter.Penumbra coils and catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2018-01438.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the internal iliac artery using penumbra coils 400 (pc400s) and a lantern delivery microcatheter (lantern).It was noted that the patient¿s anatomy was slightly tortuous.During the procedure, the physician advanced a non-penumbra catheter toward the internal iliac artery of the contralateral side from the femoral artery.While advancing a guidewire through a lantern, the physician experienced resistance approximately 30 centimeters from the tip of the lantern.Next, the physician attempted to advance a pc400 through the lantern; however, resistance was encountered approximately 30 centimeters from the tip of the lantern and the pc400 became stuck.Therefore, the physician decided to withdraw the pc400; however, the pc400 unraveled and unintentionally detached.Therefore, the lantern containing the pc400 was removed.The procedure was completed using a new lantern and additional pc400s.There was no report of an adverse effect to the patient.
|
| |
|
Search Alerts/Recalls
|
|
|