• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL; PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL; PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL 13.7
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 06/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluation; lens was returned in liquid, in a specimen cup.Visual inspection found no visible damage to the lens in liquid.(b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.7mm micl13.7, -10.5 diopter, implantable collamer lens into the patient's right eye (od) on (b)(6) 2017.On (b)(6) 2018 the lens was removed and exchanged for a shorter length lens due to excessive vault and significant reduction of irido-corneal angles.This exchange resolved the problem.Reporter stated that the cause of this event was patient related factor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7716509
Report Number2023826-2018-01095
Device Sequence Number1107749
Product Code MTA
UDI-Device Identifier00841542103572
UDI-Public00841542103572
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2018
Device Model NumberMICL 13.7
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2018
Initial Report FDA Received Date07/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age29 YR
-
-