• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. THUNDERBEAT 5 MM, 45 CM, FRONT-ACTUATED GRIP; ULTRASONIC SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. THUNDERBEAT 5 MM, 45 CM, FRONT-ACTUATED GRIP; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0545FC
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned to olympus for evaluation.The exact cause of the reported event cannot be determined at this time; however, based on similar reports, this type of teflon pad damage is most likely related to the operator's technique.The instruction manual contains several warning statements in an effort to prevent damage to the teflon pad."do not to activate output in seal and cut mode while the grasping section is closed without contacting tissue or vessel.Do not activate output while applying the probe tip to the tissue with a strong force.Do not activate output while grasping thick and hard tissues.Otherwise, various forms of damage in the probe tip and/or the tissue pad such as premature wear, breakage, deformation, exposure of metal, and/or falling inside the body cavity, and/or partial separating may occur.".
 
Event or Problem Description
Olympus was informed that during a sleeve gastrectomy procedure, the teflon pad from the grasping section of the device burnt and peeled back exposing metal.It was reported that simultaneously, a "probe damage" error message was observed on the generator.The surgeon was performing the seal/cut function when the teflon pad damage occurred.There was no bleeding reported.No device fragment fell into the patient.The intended procedure was completed with a similar device.There was no patient injury reported.Additionally, the user facility reported that the device and the connection points were inspected prior to the procedure with no anomalies found.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to report the device evaluation results, correct the device lot number and manufacture date.The device was returned to olympus for evaluation.A visual inspection was performed on the returned device and found some tissue build up.The ptfe pad (teflon pad) was inspected and found to have normal wear with a tiny area of metal exposed.There were burnt marks also noted on the top portion of the pad.The distal end of the device was inspected under a microscope and found no visual indication of damage to the probe unit.This type of teflon pad is most likely related to the operator's technique.The instruction manual contains several warning statements in an effort to prevent damage to the teflon pad."do not to activate output in seal and cut mode while the grasping section is closed without contacting tissue or vessel.Do not activate output while applying the probe tip to the tissue with a strong force.Do not activate output while grasping thick and hard tissues.Otherwise, various forms of damage in the probe tip and/or the tissue pad such as premature wear, breakage, deformation, exposure of metal, and/or falling inside the body cavity, and/or partial separating may occur.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 45 CM, FRONT-ACTUATED GRIP
Common Device Name
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
MDR Report Key7716666
Report Number2951238-2018-00426
Device Sequence Number19220869
Product Code GEI
UDI-Device Identifier04953170337543
UDI-Public04953170337543
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K111202
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTB-0545FC
Device Catalogue NumberTB-0545FC
Device Lot NumberPW304647
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/02/2018
Supplement Date Received by Manufacturer07/26/2018
Initial Report FDA Received Date07/24/2018
Supplement Report FDA Received Date08/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-