MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 36MM X 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 36MM X 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Catalog Number 71332752
Medical Device Problem Codes Loose or Intermittent Connection (1371); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Injury (2348); Joint Dislocation (2374)
Date of Event 06/21/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery was performed due to dislocation.The cause of dislocation was impingement of acetabular osteophytes.The surgeon does not fault the device.The product is not available.No more information available.
 
Additional Manufacturer Narrative
The associated r3 acetabular insert was not returned for evaluation.Our investigation including a review of complaint history on the list part revealed no additional complaints for this failure mode with the same batch number.A review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.A clinical evaluation noted that no clinical information was provided for inclusion in this investigation.The future impact to the patient beyond the revision cannot be determined. no further clinical/medical assessment is warranted at this time.Without the return of the actual product involved, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 0 DEG XLPE ACET LNR 36MM X 52MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7716749
Report Number1020279-2018-01392
Device Sequence Number16420726
Product Code MBL
UDI-Device Identifier03596010597984
UDI-Public03596010597984
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number71332752
Device Lot Number15GM05282
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/10/2018
Supplement Date Received by Manufacturer07/10/2018
Initial Report FDA Received Date07/24/2018
Supplement Report FDA Received Date10/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-