TOSOH CORPORATION TOSOH HLC-723GA ANALYZER G8 Back to Search Results
Model Number G8
Device Problems Leak/Splash (1354); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/28/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A field service engineer (fse) was dispatched to the customer's facility to address the reported event.The fse replaced the damaged waste tube and aligned the loader.The fse confirmed that the waste tube was cut short.The patient sample was not used to validate the repair.The damaged waste tubing was contaminated and disposed of at the customer site.All errors cleared and the g8 instrument was operational.No further actions were required by field service.A 13-month complaint / service history review for similar complaints was performed for the serial number (b)(4) from aware date of (b)(4) 2018, which included data from 28 may 2017 through (b)(4) 2018.There were no similar complaints identified during the searched period, which includes this event.The g8 operator's manual under chapter 6 troubleshooting is as follows: 200 area low error.Three successive results below the lower limit of the total area (50) occur.If the error message is present when sufficient volume of sample is set in the rack, the problem may be caused by an empty reagent (hemolysis & wash solution).Check the remaining volume of hemolysis & wash solution and start the assay again.The most probable cause for the reported event was the waste tank needed to be moved.
 
Event Description
A customer reported error 200 area low on the g8 instrument.The analyzer was also leaking.Technical support (ts) advised the customer to check the water heater.The water heater was full.Ts advised the customer to place the waste tank on the ground; the waste line was too short.The analyzer was not down.Customer runs patient samples on hba1c.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hba1c.There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723GA ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
ste. 101
south san francisco, CA 94080
6506368123
MDR Report Key7717114
MDR Text Key114950328
Report Number8031673-2018-00659
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 07/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA07/24/2018
Distributor Facility Aware Date06/28/2018
Device Age1 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer07/24/2018
Date Manufacturer Received06/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-