| Model Number |
SA-85 |
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/23/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A heli-fx endo anchoring system was used as an accessory device for the endovascular treatment of a type ia endoleak in a previously implanted endologix stent graft system (implanted 6 years previously).Aaa diameter measured 60mm with a proximal neck angulation of 20 degrees.50% thrombus at seal zone and 10% calcium at seal zone.During the procedure, it was reported that 8 endoanchors were implanted along with a non-mdt excluder iliac branch excluder internal iliac devices.At the end of the procedure it was reported that there was a persistent type ia endoleak present.No additional clinical sequelae were reported and the patient will be monitored by their physician.
|
| |
|
Additional Manufacturer Narrative
|
|
Failure to follow instruction for use of heli-fx endoanchors with non medtronic stent graft bifurcate device listed as not compatible in ifu.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received reported that on the (b)(6) 2018 a ct scan identified a type ia endoleak.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported by the investigator the endoleak noted on ct may be a type ii endoleak located at the lumbar artery.
|
| |
|