• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ENDO ANCHOR SYSTEM - HELI-FX AAA; ENDOVASCULAR SUTURING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ENDO ANCHOR SYSTEM - HELI-FX AAA; ENDOVASCULAR SUTURING SYSTEM Back to Search Results
Model Number SA-85
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 08/23/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A heli-fx endo anchoring system was used as an accessory device for the endovascular treatment of a type ia endoleak in a previously implanted endologix stent graft system (implanted 6 years previously).Aaa diameter measured 60mm with a proximal neck angulation of 20 degrees.50% thrombus at seal zone and 10% calcium at seal zone.During the procedure, it was reported that 8 endoanchors were implanted along with a non-mdt excluder iliac branch excluder internal iliac devices.At the end of the procedure it was reported that there was a persistent type ia endoleak present.No additional clinical sequelae were reported and the patient will be monitored by their physician.
 
Additional Manufacturer Narrative
Failure to follow instruction for use of heli-fx endoanchors with non medtronic stent graft bifurcate device listed as not compatible in ifu.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information received reported that on the (b)(6) 2018 a ct scan identified a type ia endoleak.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported by the investigator the endoleak noted on ct may be a type ii endoleak located at the lumbar artery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO ANCHOR SYSTEM - HELI-FX AAA
Common Device Name
ENDOVASCULAR SUTURING SYSTEM
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7717627
Report Number2953200-2018-01099
Device Sequence Number904025
Product Code OTD
UDI-Device Identifier00643169647695
UDI-Public00643169647695
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K140036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 10/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Model NumberSA-85
Device Catalogue NumberSA-85
Device Lot Number50207021
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/27/2018
Supplement Date Received by Manufacturer06/27/2018
08/15/2018
08/08/2018
10/18/2018
Initial Report FDA Received Date07/25/2018
Supplement Report FDA Received Date08/21/2018
09/10/2018
10/02/2018
10/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age80 YR
Patient Weight88
-
-