• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 PINN MAR NEUT 28IDX52OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 PINN MAR NEUT 28IDX52OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 121928052
Medical Device Problem Codes Fracture (1260); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Joint Disorder (2373); Limited Mobility Of The Implanted Joint (2671)
Date of Event 01/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Direct contact between the head and the acetabulum (pinnacle).That led to change in the shape of the polyethylene insert and breakage of 4 spines.Action taken was the head and the insert was changed.Primary surgery date was on (b)(6) 2017.
 
Event or Problem Description
Instability in walking.
 
Additional Manufacturer Narrative
(b)(4).Examination of the acetabular liner does find evidence to support the reported disassociation event.Product error / contribution to the event is not identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR NEUT 28IDX52OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7717675
Report Number1818910-2018-65128
Device Sequence Number16476021
Product Code LPH
UDI-Device Identifier10603295013303
UDI-Public10603295013303
Combination Product (Y/N)N
Initial Reporter CountryMO
PMA/510(K) Number
K001534
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number121928052
Device Lot NumberC41986
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2018
Initial Date Received by Manufacturer 06/26/2018
Supplement Date Received by Manufacturer08/20/2018
09/03/2018
12/12/2018
Initial Report FDA Received Date07/25/2018
Supplement Report FDA Received Date08/22/2018
09/12/2018
12/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-