| Catalog Number |
625-0T-36G |
| Medical Device Problem Code |
Noise, Audible (3273)
|
| Health Effect - Clinical Codes |
Pain (1994); Injury (2348)
|
| Date of Event |
06/27/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Surgeon commented that patient was experiencing some groin pain and squeaking while climbing stairs.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding audible noise and pain involving a trident liner was reported.The event was not confirmed.Method & results: -product evaluation and results: not performed as the device was not returned.-medical records received and evaluation: not performed as no medical records were returned for review.-product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.-complaint history review: there were no other events for the lot provided.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time as no devices and insufficient information was received by stryker orthopaedics.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
Surgeon commented that patient was experiencing some groin pain and squeaking while climbing stairs.
|
| |
|
Search Alerts/Recalls
|