| Model Number |
MMT-523NAH |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
06/29/2014
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the insulin pump had a button error alarm and high blood glucose event.The customer¿s blood glucose was unknown at the time of incident.No significant events leading to keypad anomaly were observed.Keypad anomaly troubleshooting was performed and confirmed that time is advancing.Customer also reported damage to the insulin pump battery and reservoir compartment.Customer had a high blood glucose of in the 300 mg/dl to 400 mg/dl and treated.No high blood glucose troubleshooting was performed.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is being replaced and is expected to return for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Pump received with unresponsive buttons due to moisture damage at keypad traces.No button error alarm noted.Traces of corrosion found at the electronic assembly and motor assembly per visual inspection.Unable to perform the displacement, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery and self test due to unresponsive buttons.Pump received with broken battery tube threads.
|
| |
|