• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523NAH
Medical Device Problem Code Device Difficult to Program or Calibrate (1496)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 06/29/2014
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call the insulin pump had a button error alarm and high blood glucose event.The customer¿s blood glucose was unknown at the time of incident.No significant events leading to keypad anomaly were observed.Keypad anomaly troubleshooting was performed and confirmed that time is advancing.Customer also reported damage to the insulin pump battery and reservoir compartment.Customer had a high blood glucose of in the 300 mg/dl to 400 mg/dl and treated.No high blood glucose troubleshooting was performed.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is being replaced and is expected to return for analysis.
 
Additional Manufacturer Narrative
Pump received with unresponsive buttons due to moisture damage at keypad traces.No button error alarm noted.Traces of corrosion found at the electronic assembly and motor assembly per visual inspection.Unable to perform the displacement, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery and self test due to unresponsive buttons.Pump received with broken battery tube threads.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7718163
Report Number3004209178-2018-95622
Device Sequence Number19149818
Product Code OYC
UDI-Device Identifier00643169446397
UDI-Public(01)00643169446397
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAH
Device Catalogue NumberMMT-523NAH
Device Lot NumberA8523NAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2018
Initial Date Received by Manufacturer 06/29/2018
Supplement Date Received by Manufacturer11/17/2018
Initial Report FDA Received Date07/25/2018
Supplement Report FDA Received Date11/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONCOMITANT PRODUCT: FRN-RESERVOIR
Outcome Attributed to Adverse Event Other;
Patient Age42 YR
Patient Weight135
-
-