BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLATIP MIFI OPEN-IRRIGATED ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
| Model Number |
M004PM9620N40 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.Device evaluated by mfr: the device was returned for analysis.Visual inspection revealed dried body fluid found under the ring 1 distal seal and along the distal edge of the ring 1 electrode, dried body fluid also found on the handle, main body and distal end and dried saline found on the distal end and inside the irrigation holes.Electrical tests were done and continuity checks revealed no electrical opens or shorts, as checked manually using a multi-meter and breakout box and all electrodes and thermocouple resistances measured in spec and were typical.Functional inspection was done and the steering knob and the tension control knob functioned properly on both lock and unlock positions and no abnormal resistance was felt when actuating the steering mechanism.Connected device to a maestro 4000 to measure tip temperature.During 5 minutes at room temperature (21.8c), the maestro displayed 22c and after soaking 5 minutes in a tank of 37.0c saline, the maestro displayed 37c through temperature testing.Rf ablation test was done and verified by using a maestro generator 4000, a metriq pump, and a tank of 37c saline solution.The device was found within specifications.Secondary findings revealed body fluid ingress under the distal ring 1 seal.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 19-july-2018.It was reported that the catheter had high sensor temperature display.A 7.5/110/2.5/4.5 n4 intellatip mifi open-irrigated ablation catheter was selected for an ablation procedure.During preparation, no visual issue was observed on the catheter.However, during procedure, it was noted that the catheter had high sensor temperature display when inserted into the patient.The procedure was completed with a different device.No patient complications and no injuries were reported.However, returned device analysis revealed a body fluid ingress was found under the ring 1 distal seal and along the distal edge of the ring 1 electrode.
|
| |
|
Search Alerts/Recalls
|
|
|