| Model Number |
9046 B-H |
| Medical Device Problem Codes |
Break (1069); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
| Date of Event |
07/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported there was a small hole in the reservoir.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve did not meet the requirements for leak testing due to a large tear observed in the top of the distal occluder.It is unknown how or when this damage occurred.Therefore, the patency, siphon, reflux, pressure-flow and preimplantation testing were precluded due to this damage.The instructions for use that accompany the device caution that ¿care should be taken in handling the valves as silicone has a low cut and tear resistance.¿ proteinaceous debris was observed within the interior and exterior of the valve.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that when the doctor primed the valve intra-operatively, a leak was noticed from a hole in the dome.The device had patient contact and was replaced.The patient was recovering following the procedure and there was no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|