• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUTONIX, INC LUTONIX 035 DRUG COATED BALLOON PTA CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUTONIX, INC LUTONIX 035 DRUG COATED BALLOON PTA CATHETER Back to Search Results
Model Number 9004
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Reocclusion (1985)
Date of Event 06/21/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis: the samples were not returned from the user facility; therefore, device evaluations are unable to be performed.A lot history review revealed there is no similar reocclusion complaints associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual samples were not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was not related to the study device or procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported through a clinical registry that during the index procedure, two lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion located in the left superficial femoral artery (sfa) and popliteal artery.Approximately 12 months after the index procedure, the patient¿s left sfa and popliteal vessel were reportedly reoccluded.A revascularization was performed and the hcp deemed it successful.The investigator assessed that the event was not related to the study device or procedure.The samples were discarded by the user facility and are not available for further evaluation.No adverse patient effects were reported.This is one of two products involved with the reported event and the associated manufacturer¿s report number is 3006513822-2018-00169.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 035 DRUG COATED BALLOON PTA CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
LUTONIX, INC
9409 science center dr
new hope MN 55428
MDR Report Key7719871
Report Number3006513822-2018-00168
Device Sequence Number11125860
Product Code ONU
UDI-Device Identifier00801741088766
UDI-Public(01)00801741088766
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,stu
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/04/2018
Device Model Number9004
Device Catalogue NumberLX351306150
Device Lot NumberLUAR0094
Was Device Available for Evaluation? No
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight74
-
-