Consumer reported complaint for low blood glucose test results.Daughter is calling on behalf of the customer.The customer's expected non-fasting blood glucose test result range is 200 - 300 mg/dl; customer normally tests 1 hour after eating in the evenings.At the time of the call, the customer felt well and did not report any symptoms.Daughter stated that the night before customer had performed a blood test fasting and had obtained a result of 282 mg/dl.Daughter stated based on the result of 282 mg/dl customer had administered 8 units of insulin.When customer tested again non-fasting he had obtained blood glucose test results of 45, 40 and 33 mg/dl.Due to the low results, daughter had given him applesauce.Customer was also not coherent and had no strength.Daughter had called ems.Ems arrived and customer's blood glucose test result with their meter was 35 mg/dl non-fasting.Ems had administered sugar to the customer.Customer was concerned that his glucose should not have dropped so low so fast and was concerned with the non-fasting low results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is stored according to specification in the living room.Test strip lot manufacturer's expiration date is 04/30/2019 and open vial date is three weeks.The meter memory was reviewed for previous test result history: result 1 :171mg/dldate: (b)(6) 2018 time:732am fasting; result 2 :113mg/dldate: (b)(6) 2018 time:1043pm non- fasting; result 3 :33mg/dldate: (b)(6) 2018 time:1025pm non- fasting; result 4 :40mg/dldate: (b)(6) 2018 time10:12pm non- fasting; result 5 :45 mg/d date: (b)(6) 2018 time10:07pm non- fasting; result 6: 282mg/dl date: (b)(6) 2018 time:744pm fasting.
|